Kingon P2 Range - Alert & Error Codes

Kingon P2 Range - Alert & Error Codes

Kingon P2 Range - Alert & Error Codes

This page helps you understand the alerts and error codes on your Kingon P2-E Oxygen Concentrator, so you can quickly identify what the device is communicating and what action, if any, is required.

P2-E
P2-E6 / E7
P2-E
P2-E6 / E7


Power / Startup / Temperature - Alarm codes:  E01 – E07  

Alarm CodeDescriptionAlarm ConditionSystem Response
E01Power Supply FailSystem voltage below 10.5VAutomatic shutdown after 10 seconds
E02System Startup FailOxygen concentration below 87% after startupAlarm only
E03Gas Supply FailFlow or concentration below normal operating rangeAlarm only
E04Battery OverheatingBattery temperature above 65°CAutomatic shutdown after 10 seconds
E05System OverheatingSystem temperature above 60°CAutomatic shutdown after 10 seconds
E06Battery Temperature LowBattery temperature below 0°CAlarm only
E07System Temperature LowSystem temperature below 0°CAlarm only

Battery / Mechanical Failures -  Alarm codes:  E08 – E0C 

Alarm CodeDescriptionAlarm ConditionSystem Response
E08Battery DepletedRemaining state of charge below 5% without power adapterAutomatic shutdown after 10 seconds
E09Cooling Fan FailureCooling fan malfunction detectedAutomatic shutdown after 10 seconds
E0AValve Check FailureValve switch malfunction detectedAutomatic shutdown after 10 seconds
E0BCompressor FailureCompressor malfunction detectedAutomatic shutdown after 10 seconds
E0CSieve Bed FailureSieve bed fault or invalid operation detectedAutomatic shutdown after 10 seconds

Oxygen / Gas Delivery -  Alarm codes:  E0D – E012 

Alarm CodeDescriptionAlarm ConditionSystem Response
E0DTank Pressure FailureTank pressure fault detectedAlarm only
E0EGas ObstructionPipe or nasal cannula blockedAlarm only
E0FGas Delivery FailureNo gas delivery detected after injectionAlarm only
E10Oxygen Sensor FailureOxygen sensor malfunction detectedAlarm only
E11Breath Sensor FailureBreath sensor fault detected or cannula kinkedAlarm only
E12Low Oxygen ConcentrationOxygen concentration below 50% for more than 10 minutesAlarm only

Battery / Oxygen Warnings -  Alarm codes:  E13 – E17 

Alarm CodeDescriptionAlarm ConditionSystem Response
E13Low BatteryRemaining state of charge between 5% and 20%Alarm only
E14Low Oxygen ConcentrationOxygen concentration below 87% for more than 10 minutesAlarm only
E15Absence of BreathNo breath detected for more than 60 secondsAlarm only
E16Low Input VoltagePower adapter input voltage below 17VSwitch to battery power
E17Battery ExhaustedBattery exceeds 500 charge cycles or health below 50%Alarm only